COLOR CALIBRATION AND COLOR MANAGED WORKFLOW FOR MEDICAL DISPLAYS





             




Introduction

This document discusses methods for consistent color and grayscale presentation for digital color displays to be used in medical interpretation. There are advantages in using color displays over grayscale displays. Users can apply different modalities to the same display. Furthermore it is easier to use the application software on a color display as the navigation on a color display is more clear thus increasing users productivity. Above all, color displays are required in pathology, telemedicine and ophthalmology, because of the inherent color in these medical imaging methods.

Due to technical developments in display manufacture, more displays are and will be capable to display colors. LCD manufacturing costs were reduced drastically due to the high output volume in the TV market. Cold cathode fluorescent lamps (CCFLs) are being replaced by white LED, RGB LED backlight and OLED (Organic Light Emitting Diodes). One of the first real 10-bit format or panels comes from the consumer market in the formats of 16:9 and 16:10 with internal LUT's of up t0 14 LUT. These technologies provide display manufacturers with the capability to produce displays with bigger gamut, higher luminance, lower heat emission and lower price.


Why are there now more color applications

Large numbers of image rendering devices will allow generation of even larger numbers of clinical images.

Some of those images are displayed for making a diagnosis, reviewing live in conferences or treatment (endoscopy). Some images are true RGB color images and some are pseudo-color images (with COLOR PALETTE Photometric interpretation).


Why are color and luminance consistency important

Color and grayscale consistency makes sure that the same color is produced and reproduced across different output and input devices. These devices can be displays, printers, projectors, cameras or microscopes. This high accuracy in color and luminance reproduction is required due to the high sensitivity of the human observer with respect to luminance differences as well as color differences. Consequently images are required with equidistant color and gray levels, consistent and correct color-reproduction over the entire work-flow process in order to enable the observer to distinguish all color and gray shades that exist in the color image file. 


The need for Color Management in medical imaging

Managing color is a complex procedure and was addressed already in the pre-press industry. 

The International Color Consortium has developed a system to manage colors using a so called ICC profile and a CMM (Color Management Module). The ICC profile does characterise the imaging device and the CMM is a software algorithm that adjusts the numerical values, that get sent to or are received from different devices so that the perceived color they produce remains consistent. 




Figure 1: Color management system8





Figure 2: QUBYX PerfectLum Display calibration and verification software



QUBYX

QUBYX is a technology contributor to the Medical Imaging and Color Management Industries with various professional Calibration and Visualization solutions since 1996. QUBYX develops software solutions that are used in and for computer displays. Our solutions include OSD replacement software, Display Management systems, Color Calibrations and Verification software for the Medical imaging use. QUBYX has an extended knowledge in digital color technologies. Our knowledge extends to handling, calculating, mapping from different color spaces and creating ICC profiles. For More information, visit http://www.qubyx.com




QUBYX is working on a system to calibrate,verify and evaluate the accuracy of color and luminance calibration of medical displays and prints.




Figure 3: Calibration and verification



Verification includes deviations towards the target luminance response function and the achieved color reproduction.



Figure 4: Verification of JND (Just Noticeable Differences)





Figure 5: Verification of luminance response as described in AAPM TG182


Figure 6: Verification of Delta E towards the reference color-temperature over DDL (Digital Driving Level)




Figure 7: Color Image Reproduction Process for ICC Managed Color Medical Imaging



EXISTING STANDARDS

DICOM GSDF:
Part 14 of the DICOM standard3 describes a function on how an output device like a printer or display shall reproduce luminance output over driving level.


Figure 8: The Grayscale Standard Display Function presented as logarithm-of-Luminance versus JND-Index


Limitations of existing NEMA DICOM Part 14: 
  • applies only to grayscale images 
  • only grayscale consistency over devices 
  • no pseudo colors 

The DICOM Supplement 100 "Color Softcopy Presentation State"5 fills this void and guesses how color should be managed and gives a guideline on how to implement a color-reproduction pipeline and device-independent color spaces.

Basic recommandations of the supplement 100 are:

  • Use existing industry standard of ICC profiles 
  • Use the device-independent Profile Connection Space (PCS), CIEXYZ or CIELAB, as defined by the ICC 
  • The device depended RGB color space shall be sRGB 
  • The rendering intent is fixed to "perceptual" 
  • Space LUTs should be represented as 16 bit values, using LUT 16 type Tag, for greater precision. 
  • The chromaticAdaptationTag should be not set if the actual illumination source is D50. 
  • Color image files should have embedded ICC profiles 

CIE L*

The CIEL L* function is based on the CIE Lab colorspace11 and is reproducing the sensibility of the human observer9 on luminance and color shades.


Figure 9: The CIE L* function


ICC color profile standard

The International Color Consortium promotes a standard for color Specification, an ICC profile, that can be associated with images or devices3 . The profile permits interpretation of numeric values in a multi-channel image as true colors. An ICC profile is a set of data that characterizes a color input or output device, or a color space, according to standards promulgated by the International Color Consortium (ICC). Profiles describe the color attributes of a particular device or viewing requirement by defining a mapping between the device source or target color space and a profile connection space (PCS). This PCS is either CIELAB (L*a*b*) or CIEXYZ. Mappings may be specified using tables, to which interpolation is applied, or through a series of parameters for transformations.

To profile the monitor, again a color sensor is attached flat to the display’s surface. Profiling software sends color and grayscale shades to the display and compares the values actually sent against readings from the calibration device, creating a matrix for the display.

Depending on the calibration software, different versions of ICC profiles are created like version2 or version4 and with different tags either CIELAB (L*a*b*) or CIEXYZ. Mappings may be specified using tables, to which interpolation is applied, or through a series of parameters for transformations.

Every device that captures or reproduce color can have its own profile.


Depending on the profiling application and the device, different TAG´s are embedded into an ICC profile.
red/green/blueMatrixColumnTag´s
red/green/blueTRCTag´s
AToB0Tag, AToB1Tag, AToB2Tag, BToA0Tag, BToA1Tag, BToA2Tag



Many named color profiles exist: sRGB and Adobe RGB are in common use as the RGB workspace.


Figure 10: CIE 1931 Chromaticity diagram6. 
The range of colors perceptible to the human visual system is represented by the color gamut shown here in CIE 1931 x,y coordinates.



Methods of Display and Printer calibration and profiling


Calibration

The aim of color calibration is to measure and adjust the color response of a device (input or output) to establish a known relationship to a standard Luminance response like DICOM GSDF or CIE L* and to a standard white point like D65.

Color calibration is a requirement for all devices taking an active part of a color/luminance managed workflow. A measurement device is needed to measured colors and luminance of the Displays and Printers. The measured colors are described in standard color spaces like CIE Lab or CIE XYZ.


Figure 11: CIE xy color space with sRGB profile Gamut view and with colortemperature curve of the blackbody


Display calibration

For calibrating the monitor a color sensor is attached flat to the display's surface, shielded from all ambient light. The calibration software sends a series of color and grayscale shades to the display and compares the values that were actually sent against the readings from the calibration device. This establishes the current offsets in color display. Depending on the calibration software and type of monitor used, the software will load this correction table into the Graphics Board or into the Display.



Figure 12: Luminance response before calibration (Red) CIE L* Target (Green) and result after calibration (Blue)




Figure 13: Gamma correction that was calculated and applied to reach the target color and luminance response



Display profiling

For profiling of the monitor again a color sensor is attached flat to the display's surface. The profiling software sends a series of color and grayscale shades to the display and compares the values that were actually sent against the readings from the calibration device. This creates a matrix for the display. Depending on the calibration software different versions of ICC profiles are created like version2 or version4 and with different Tags.



Figure 14: ICC Profile map




Figure 15: The Gamut of an Display ICC profile in 3D view


Figure 16: Once the Display ICC profile is created it has to be selected as the default ICC profile in the Operating system.


Printer calibration and profiling

For calibrating the printer a spectrophotometer or densitometer is used to measure the ink on the paper. The calibration software create a document with a series of color and grayscale shades and prints it. The print is read with the spectrophotometer or densitometer. The software compares the values that were actually sent against the readings from the measurement device. This establishes the current offsets for the printer. Depending on the calibration software and type of printer used, the software will load a correction table into the print server (RIP) or creates an ICC Profile only. The calibration needs to be redone once paper or ink is changed inside the printer.




Figure 17: Color or grayscale printers can be calibrate to DICOM GSDF




Figure 18: A spectrophotometer or densitometer is used to calibrate and profile a printer



Color management workflow for different modalities


Color Images





Grayscale Images





Pseudocolor Images

Figure 19 (all 4 images together)



The correction LUT (Look Up Table) of the calibration is applied in the display, graphicboard, camera, printer, microscope. The LUT is adding an offset to the color input values and create a new output. The ICC profiles needs to be selected in the system and/or inside the images application or printing and camera driver.

Rendering intent


When the gamut of source color space exceeds that of the destination, saturated colors are liable to become clipped (inaccurately represented), or more formally burned. The color management module can deal with this problem in several ways. The ICC specification includes four different rendering intents: absolute colorimetric, relative colorimetric, perceptual, and saturation.

Absolute colorimetric

If the output device has a much larger gamut than the source profile, i.e., all the colors in the source can be represented in the output, using the absolute colorimetry rendering intent would "ideally" (ignoring noise, precision, etc.) give an exact output of the specified CIELAB values. Perceptually, the colors may appear incorrect, but instrument measurements of the resulting output would match the source. Colors outside of the proof print system's possible color are mapped to the boundary of the color gamut. Absolute colorimetry is useful to get an exact specified color for example like in pathology or pseudo colors.

Figure 20


Relative colorimetric

The goal in relative colorimetry is to be truthful to the specified color, with only a correction for the visual adaptation due to the media (paper color in the case of a printer) or white point in the case of a display. Relative colorimetry is useful in proofing applications, since you are using it to get an idea of how an image on one device will appear on a different device.


Perceptual and Saturation

Perceptual rendering is recommended for color separation. In practice most used is relative or perceptual intent, as for natural images, absolute causes color cast, while saturation produces unnatural colors. Relative intent handles out-of-gamut colors by clipping (burning) these colors to the edge of the gamut, leaving in-gamut colors unchanged, while perceptual intent smoothly moves out-of-gamut colors into gamut, preserving gradations, but distorts in-gamut colors in the process. If an entire image is in-gamut, relative is perfect, but when there are out of gamut colors, which is more preferable depends on a case-by-case basis. Saturation intent is most useful in charts and diagrams, where there is a discrete palette of colors which one wishes to have saturated, but where the specific hue is less important.


Figure 21


Figure 22


Display White and Black level and ambient light

Display white and black level must be constant over time as well as the ambient light. The ambient light should be low and no light reflection should be visible on the display panel.

Still unexplored and missing items.

  • PACS applications and DICOM viewers miss ICC profile support. Without ICC profile from the imaging application the color correction will not be applied. Software developers need to integrate the ICC support. 
  • The DICOM Supplement 100 "Color Softcopy Presentation State” does not specify the luminance response a color display shall be calibrated. Should the luminance response be GSDF or CIE L* or gamma? Should color be corrected on a color display where grayscale images are displayed? Should an ICC profile be created for color displays where only grayscale images are displayed? 
  • The DICOM Supplement 100 "Color Softcopy Presentation State” fixed the rendering intend to “Perceptual”. Perceptual will make sure that the perception of the entire images stays intact. But for applications where the exact color is more important than the global perception like in applications like Endoscopy, Microscopy or in case of pseudo-colors the “absolute” rendering intent would be here more appropriate. 
  • The DICOM Supplement 100 "Color Softcopy Presentation State” does not specify a white point or any standard illuminant10 to calibrate too. Without a specific white point there can not be consistency from display to display. 
  • The DICOM Supplement 100 "Color Softcopy Presentation State” does not specify how to verify and evaluate the colorcalibration like it is done for DICOM GSDF. Adjustment and verification of the white point over the entire dynamic range is not described in DICOM 

References:


1) H. Roehrig et al:”Color calibration and color-managed displays: Does the calibration method matter?” Proc SPIE Vol 7627; pp 76270K1-13. 
2) AAPM TG18 report http://deckard.mc.duke.edu/ ~ samei/tg18 
3) Digital Imaging and Communications in Medicine (DICOM) 
4) Part 14: Grayscale Standard Display Function http://medical.nema.org/dicom/2004/04_14PU.PDF
5) Supplement 100: Color Softcopy Presentation Stateftp://medical.nema.org/medical/dicom/final/sup100_ft.pdf 
6) CIE Space http://www.fho-emden.de/ ~ hoffmann/ciexyz29082000.pdf 
7) ICC profiles http://color.org/ 
8) Colormanangement http://www.kleocolor.com
9) Colorimetry -- Part 1: CIE standard colorimetric observers ISO 11664-1:2007 (CIE S 014-1/E:2006) 
10) Colorimetry -- Part 2: CIE standard illuminants ISO 11664-2:2007 (CIE S 014-2/E:2006)
11) Colorimetry -- Part 4: CIE 1976 L*a*b* Colour space ISO 11664-4:2008 (CIE S 014-4/E:2007)

Friday, October 25, 2013

CALIBRATING COLOR MEDICAL DISPLAYS: HOW TO DO IT RIGHT?


Color medical displays are in, and they are in for good. While with grayscale display calibration things are very clear due to DICOM Part 14 GSDF standard, requirements to color calibration of medical monitors are still somewhat uncertain.

Calibrating a display to read grayscale images normally means adjusting the luminance response curve. Luminance response, also known as gamma or transfer function, is a curve that describes the luminance output of an image producing device like a display or a printer.

DICOM Part 14 recommends calibrating a display to the Barten luminance response curve, also known as the DICOM curve. The DICOM curve is widely used and proved to work very well for grayscale images. It ensures that all gray shades are displayed with equal distances for the human eye.

Here’s what just noticeable differences mean on practice:


 










* Inconsistent image display. The lump visible on the left is almost invisible on the right.













* Same images. The image on the left is read on a high brightness, but uncalibrated display. Not all JND’s are visible. The image on the right is displayed by a low luminance, but calibrated monitor. More JND’s are visible.

Since DI
COM Part 14 was originally developed for grayscale displays, it may not be a universal solution for color medical monitors as well. So, what luminance response do you calibrate a color medical display to if you use it to read grayscale images? Or color and pseudo color images? To answer these questions, let’s take a look at the available options for luminance response calibration.


What Are the Options for Luminance Response Calibration?

Barten/DICOM

The DICOM Grayscale Standard Display Function specifies exactly what luminance or density level should be produced for a certain input value, based on the Barten curve, which maps the values into a range that is perceptually linear. This means that input values are mapped into a space that is perceived as linear by a human observer.

With the luminance response adjusted to the DICOM curve, the display presents more just noticeable luminance differences to the viewer, and achieve similarity in grayscale reproduction for various displays.

DICOM Part 14 calibration makes sure all gray levels are produced in equal distance for the standard observer. Obviously, this type of luminance response calibration was introduced for grayscale images.

Gamma

Gamma (gamma correction, gamma encoding)  a nonlinear operation used to code and decode luminance. Gamma correction is defined by a power-law expression.

Gamma was originally developed to compensate for the characteristics of CRT displays.  In CRT displays, the light intensity varies nonlinearly with the electron-gun voltage. Altering the input signal by gamma compression can cancel this nonlinearity, such that the output picture has the intended luminance. So, Gamma correction is a good option for a CRT monitor, or to reproduce a CRT display’s image.

CIE L*

CIE luminosity function is established by the Commission Internationale de l'Éclairage (CIE) to describe the average sensitivity of the human eye to lightness. So in a certain way it is similar to the DICOM curve, but CIE L* does not only describe the equal distance in gray shades, but also the equidistant reproduction of colors and color shades. Calibrating color displays to CIE L* is currently recommended by the DIN 6868-57 standard (a German standard, discussing performance of medical displays).

Rec. 709

This luminance function is part of the Rec. 709 standard, developed for high-definition TV.
Thus this standard is highly adapted to pictures in movement. High definition endoscopy devices should follow REC 709.

EPD

EPD aim curve is part of a standard for geospatial imaging displays by National Geospatial Intelligence Agency (NGA). According to this standard, luminance response must match the EPD curve to ensure equal probability of detection for every shade of gray and every detail.

sRGB

sRGB is a standard developed cooperatively by HP and Microsoft  for the use on monitors, printers and the Internet.
sRGB uses a transfer function typical for CRT displays. This specification allowed sRGB to be directly displayed on typical CRT monitors of the time, a factor which greatly aided its acceptance.
The sRGB gamma can not be expressed as a single numerical value. The overall gamma is approximately 2.2, consisting of a linear (gamma 1.0) section near black, and a non-linear section elsewhere involving a 2.4 exponent and a gamma (slope of log output versus log input) changing from 1.0 through about 2.3.

This standard  color space is widely used as a default color space for various image reproduction devices, like displays, printers, or scanners. sRGB uses luminance response, typical of CRT displays and is close to Gamma 2.4.

CIECAM02

Here luminance, among other characteristics of a monitor, is adapted according to the CIECAM02 color appearance model, developed by the CIE.


NTSC

National Television System Committee is an analog television standard that defines certain luminance response among other characteristics of a TV.


PAL/SECAM

This is a standard color encoding system for analog television, defining luminance response among other characteristics.



Why Is Calibrating Only Luminance not Enough?

This image shows how two monitors, calibrated to the same luminance level, display the same shade of gray. The luminance is the same on both images, but for the human eye the grays look totally different. This difference can be disturbing and interfere with the way the viewer perceives and interprets an image.



* The difference you see is caused by the different color temperature of the grays, 
so even if you use the color display for reading grayscale images only, calibrating luminance to DICOM may prove insufficient. For the optimal performance of a color medical display, you need to calibrate both luminance response (Gamma) and color temperature of the monitor.

































  
* All these levels of gray have the same luminance inside one row. 
But the human eye  would eventually see one gray shade as brighter or darker than another. That is why color calibration over the entire  luminance range is so important, and the highest accuracy is required with a delta Eab of at least < 2.



What Is the Right Choice?

As you see, there is no universal solution for any display and any purpose. If you use a color display for viewing grayscale images, the best option is to calibrate the display to DICOM GSDF and calibrate the color temperature, too.

For reading color and pseudo color images, we recommend calibrating the monitor to CIE L* and calibrating the color temperature. The recommended color temperature is D65: it corresponds to a Standard Norm light from the CIE (Commision Internationale D'Eclairage).

Finally, if you need to view moving images, it is best to calibrate your monitor to a video-related standard. We recommend calibrating color temperature, and calibrating luminance response to Rec.709.


Wednesday, July 17, 2013

MEDICAL IMAGING ON MAC: HOW MAC'S COLOR MANAGEMENT CAN GET YOUR DISPLAY OFF CALIBRATION

Immensely popular with consumers, Mac machines are gradually - and quite predictably - winning the medical imaging market. More and more Mac machines are used for diagnostic and reviewing purposes, especially as home workstations. The open source DICOM viewer Osirix enjoys great success in the medical world, and so does its FDA-cleared and CE-certified version, offered by Pixmeo.

But the thing is the sophisticated color management system Macs are equipped with can be quite tricky for a medical display. It offers many advantages for correct color and luminance reproduction, but can also present a risk, if set up incorrectly.

How Does Mac's Color Management Work?


Mac OS uses ICC profiles to apply a correction LUT (generally referred to as Gamma) to the desktop, but also applies 3D LUT’s inside the imaging application. Color management can not be deactivated on Mac, and a correction will always be applied both to the desktop and inside the imaging application.